SPECIAL INSTRUCTIONS

Mixed frequency stimulation (P4)
Mixed frequency stimulation is a combination of high (80 Hz) and low (2 Hz) frequency stimulation
with separate amplitude settings.
1. First set the amplitude for 80 Hz stimulation when the upper part of the high/low frequency symbol
is flashing
2. When the bottom part of the symbol flashes for the first time, the amplitude for 2 Hz stimulation will
start at the same level as is set for 80 Hz. Adjust the amplitude until you get visible muscle contrac-
tions.
The period between the frequency changes is preset to 3 seconds.
See chapter FREQUENTLY ASKED QUESTIONS for further information on mixed frequency stimulation.
Intermittent stimulation + hand switch (P9–P23)
Programs 9–23 are intermittent stimulation programs
with rest time between the muscle contractions (work
time), see figure. The contraction gradually increases
during ramp-up time and reaches maximum contrac-
tion during work time. During ramp-down time, the
stimulation gradually decreases until the rest time
starts. The rest time can be without or with stimulation (Active Rest). In intermittent stimulation pro-
grams the work/rest symbol
ing work time and the bottom part during rest time.
Programs P9–P16 and P19–P23 can be used together with the hand switch to manually control the
duration of the work time and rest time. Press the hand switch button during rest time for ramp-up
time and contraction. The duration of the work time is preset but can be shortened by pressing the
hand switch button. The rest time lasts until the hand switch button is pressed again.
See chapter FREQUENTLY ASKED QUESTIONS for further information on using the hand switch.
Intermittent stimulation with Active Rest (P19–P20)
In programs with Active Rest, the stimulation is active also during the rest time, which means that you
have to perform two amplitude settings for each channel.
Amplitude setting for contractions (work time): When the upper part of the work/rest symbol
is flashing, increase the amplitude gradually until you get muscle contractions without pain.
Amplitude setting for Active Rest: When the bottom part of the work/rest symbol
increase the amplitude gradually until you get muscle vibrations.
See chapter FREQUENTLY ASKED QUESTIONS for further information on Active Rest.
1
1. Ramp-up time
2. Work time
is shown on the display. The upper part of the symbol is flashing dur-
2
2
3
1
4
3. Ramp-down time
4. Rest time
is flashing,
15
3