Level 1
Level 2
SIMULTANEOUS
Level 3
Conventional
(C)
Pulse duration
Preset: 180 µs
Range: 50–400 µs
Frequency work
Preset: 80 Hz
Range: 1–120 Hz
Frequency rest*
Preset: 0 Hz
Range: 0–10 Hz
Work time
Preset: 3 s
Range: 1–30 s
Rest time
Preset: 10 s
Range: 5–60 s
Ramp-up time work
Preset: 2 s
Range: 0–10 s* **
Ramp-down time work
Preset: 2 s
Range: 0–10 s* **
Time
Preset: – – min.
Range: 1-99, – – min.
INTERMITTENT
(I)
(SI)
Frequency
modulation
(FM)
Pulse duration
Preset: 180 µs
Range: 50–400 µs
Frequency work 1
Preset: 5 Hz
Range: 2–120 Hz
Frequency work 2
Preset: 15 Hz
Range: 2–120 Hz
Frequency rest*
Preset: 0 Hz
Range: 0–10 Hz
Work time
Preset: 5 s
Range: 5–30 s
Rest time
Preset: 10 s
Range: 5–60 s
Ramp-up time work
Preset: 2 s
Range: 0–10 s* **
Ramp-down time work
Preset: 2 s
Range: 0–10 s* **
Time
Preset: – – min.
Range: 1-99, – – min.
ALTERNATING
(ALT)
Pulse duration
Preset: 180 µs
Range: 50–400 µs
Frequency work
Preset: 80 Hz
Range: 1–120 Hz
Work time
Preset: 3 s
Range: 1–30 s
Ramp-up time work
Preset: 2 s
Range: 0–10 s***
Ramp-down time work
Preset: 2 s
Range: 0–10 s* **
Rest time**
Preset: 10 s
Range: 5–60 s
Time
Preset: – – min.
Range: 1-99, – – min.
* Frequency between 1 and 5 Hz a Burst stimulation
Other frequency a Conventional stimulation
** Rest time must be greater than
Work time + Ramp-up time work + Ramp-down time
work
*** 0.5 s is shown as 05 on the display.