7 Conditioning
44
Operating Instructions Midwest E Pro / E Plus
1. Brush the instrument under running water (< 38 °C, < 100 °F, at
least drinking water quality) and good lighting (min. 500 lux) and
color rendering index (min. 80 Ra) until no more dirt can be seen, for
at least 10 seconds.
2. Flush the drive channels with spray.
3. Clean and disinfect spray channels with suitable agents and
adapters according to manufacturer's instructions.
4. Blow spray channels with suitable adapters out with 2.5 - 3 bar until
no more moisture displaces, but at least 10 seconds.
5. Conduct thermal disinfection or unwrapped steam sterilization.
6. Maintain mechanical parts manually [ → 45].
7. Maintain the push button chuck manually [ → 46].
8. Pack the instrument in packaging material suitable for sterilization
and storage. e.g. paper/plastic composite packaging.
9. Conduct sterilization [ → 48].