Local display and operation
1.
Go to Local Display and Operation > Local Operation:
Lock
options
HW Write
Parameter changes using SIMATIC PDM and
Protection
settings via local operation are both disabled.
Independent of the other lock functions.
Write locking Prevents parameter changes via the bus. Local
operation is possible.
Local
When disabled, no access is possible via the
Operation
keypad. After a communication failure, local
operation is automatically enabled 30 s later. Once
communication is reestablished, the original
setting for Local Operation is restored.
Locking functions can be combined:
HW Write Protection
Off
On
Off
Off
Off
2.
Go to Local Display and Operation > Display source:
select either Output Value, Primary Value, Sensor Value, or Electronics Temperature.
3.
Go to Local Display and Operation > Local Status Text:
select a language, or numeric option.

Sensor Test

Open the Device Menu to find and activate Sensor Test via PDM, and to see the results:
(test successful, or test failed).
Error Messages and References: PROFIBUS PA
For more detail, please consult the PROFIBUS PA User and Installation Guideline
Note:
(order number 2.092), available for download from www.profibus.com.
Page EN-16
Pointek CLS200/300 (Digital) – QUICK START MANUAL
All manuals and user guides at all-guides.com
Effect
Write locking
Off
On or Off
Off
On
On
Turn on/off
Keypad
Mode 10
SIMATIC PDM
(via Device Menu)
SIMATIC PDM
Local Operation
enabled
enabled or disabled
disabled
disabled
enabled
Digital
display
L
Lc
LA
Digital display
L
LA
LL
Lc
7ML19985XA82