Process Connections; Ambient Temperature - Siemens CLS200 Manual De Inicio Rápido

Ocultar thumbs Ver también para CLS200:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 75
Pointek CLS200/300 (Digital)
Note: Pointek CLS200/300 is to be used only in the manner outlined in this manual, otherwise
protection provided by the equipment may be impaired.
This product is intended for use in industrial areas. Operation of this equipment in a
residential area may cause interference to several frequency based communications.
Pointek CLS200/300 is a versatile capacitance
switch with a high level of chemical
resistance; ideal for level detection of
interfaces, solids, liquids, slurries, and foam,
and for simple pump control.
Approvals (verify against
device nameplate)
• CE, CSA
, FM, ATEX
C/US
• Vlarem II, WHG
• Lloyd's Register of Shipping, categories
ENV1, ENV2, and ENV5
Note: The use of approved watertight conduit hubs/glands is required for Type 4 / NEMA 4,
Type 6 / NEMA 6, IP68 (outdoor applications). For CE requirements, the use of EMC-rated
cable entries is required.

Process connections

Compact (std.) configuration
Sanitary configuration
Cable and slide coupling
configuration

Ambient temperature

• General applications
- local display
- storage temperature
• In potentially explosive atmospheres
Process Conditions
Please see full Operating Instructions for Process Pressure/Temperature De-
Note:
rating Curves.
• relative dielectric constant (
• CLS200 temperature
- without thermal isolator -40 to +85 °C (-40 to +185 °F)
- with thermal isolator
1.
At process connection.
7ML19985XA82
Pointek CLS200/300 (Digital) – QUICK START MANUAL
All manuals and user guides at all-guides.com
R ¾", 1", 1 1/4", 1 ½" BSPT; ¾", 1", 1 1/4", 1 ½" NPT; G ¾", 1",
1 ½" BSPP
1", 1 ½", 2", 2 ½"and 3" tri-clamp
R ¾", 1", 1 1/4", 1 ½" BSPT; ¾", 1", 1 1/4", 1 ½" NPT (Taper);
G ¾", 1", 1 ½" BSPP
-40 to +85 °C (-40 to +185 °F)
-30 to +85 °C (-22 to +185 °F)
-40 to +85 °C (-40 to +185 °F)
(check temperature class shown on device
nameplate)
1.5 minimum
r
1
:
-40 to +125 °C (-40 to +257 °F)
Digital version
lid with window
integral digital display
optional M12 plug
available
device
nameplate
Page EN-3

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Cls300Pointek cls200Pointek cls300