coltene/whaledent Coltolux LED Manual Del Usuario página 4

Ocultar thumbs Ver también para Coltolux LED:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 16
86768C Coltolux LED OG 7x8_Intl.qxd
Owner's Guide
C. If the On/Off button on the
handpiece does not activate the
curing light and the green light
above the button is illuminated,
unit requires service. Contact your
dealer or Coltène/Whaledent Inc.
D. If your light registers a reading
below 300 mW/cm
2
on a light-
meter, take the following steps:
• Check the light lens.
The light lens should be
checked daily. Replace lens with
Coltène/Whaledent Inc. Light Lens
Cat. No. C7974 if material has
cured to the lens.
• If the intensity of the light output
is still under 300 mW/cm
the curing time according to
the material's manufacturer
recommendation until the
®
Coltolux
LED light can be
serviced. Contact your dealer or
Coltène/Whaledent Inc. for service.
VII. REPLACING THE
LIGHT LENS
IMPORTANT: The Coltolux
Light Lens is easily scratched and
must be handled with care at all times.
CAUTION: Only use
Coltène/Whaledent's Light Lens
Cat. No. C7974. Do not use other
manufacturer's light lens with this
product.
CAUTION: Remove the
handpiece from the charger base
before attempting to replace the
light lens. Do not activate the
handpiece.
• Remove the light lens by rotating
the lens counterclockwise until
the lens disengages from the
handpiece.
CAUTION: Do not touch the LED
under the light lens as hand oil
contaminates the LED and can
compromise performance.
12/1/04
11:11 AM
VIII. INSPECTING THE
LIGHT LENS
• Inspect the light lens before each
use to insure that restorative
material has not accidentally
cured to the lens. Replace lens
with Coltène/Whaledent Inc. Light
Lens Cat. No. C7974 if material
has cured to the lens.
CAUTION: Never look directly
into the light emitted from the
light lens or reflected from tooth
or other surfaces.
IX. INSPECTING THE
CHARGER BASE POWER
CORD
2
, extend
• Periodically inspect the charger
base power cord for frayed
areas or broken insulation.
X. CLEANING
• All external components may be
cleaned by wiping with a soft
cloth moistened with isopropyl
alcohol.
IMPORTANT: Do not use petroleum-
®
LED
based solvents, iodophors, or
phenolic-based products on the
handpiece or charger base.
(Iodophors and phenolics can stain
the surface of the unit and damage
the plastic housing.)
XI. INFECTION CONTROL
• To comply with infection control
procedures the light shield is to be
placed on the handpiece first, then
slide the barrier sleeve over the
handpiece and shield. A new
barrier sleeve should be used for
each patient.
XII. SPECIFICATIONS
A. Electrical Safety
®
Coltolux
LED is a tested and
listed device under UL
and CSA standards.
Page 5
This product fulfills the
requirements concerning
electromagnetic compatibility in
accordance with EMC directive
93/42/EEC.
B. Dimensions (approximate)
Handpiece
Length:
Probe Dia:
Grip Dia:
Charger Base
Height:
Width:
Depth:
C. Operating Voltage/Power
Power Supply
Input:100 – 240 VAC, 50 – 60 Hz
Output: 5 VDC, 6 watts
Class:
Handpiece
Operating Voltage: 4.2 – 3.0 VDC
D. Environmental Conditions
®
Coltolux
operate in ambient temperatures
from 50°F – 104°F (10°C – 40°C),
relative humidity 80% at 77°F (25°C).
CAUTION: To maximize patient
comfort, do not operate
handpiece for more than one
minute continuously if the
ambient temperature is above
30° C.
E. Classifications & Symbols
Coltolux
IEC60601-1 Class II, Type BF, for
protection against electric shock.
Page 5
22.1 cm / 8.7"
1.22 cm /.48"
2.5 cm / 1.00"
7.4 cm / 2.9"
6.4 cm / 2.5"
8.9 cm / 3.5"
II
LED is designed to
®
LED is classified as
IEC Symbol
Publication: 878-02-03
Type BF Equipment