Message
The standard value is out
of the measurement
range.
The standard value is too
low.
The standard value is too
high.
Verification failed.
Instrument update
Copy from USB Memory
failed
Instrument update file is
missing
Instrument update file is
corrupt
Not enough memory to
update the instrument
USB memory is not
connected.
Read/Write to USB flash drive
Cannot write to USB
memory
Cannot read from USB
memory
Restore backup
No instrument backup is
available.
Not able to restore the
backup
22 English
Use the correct calibration standards. Invert the standards. Make sure
to measure the standards in ascending order.
The wrong calibration standard is in the vial compartment. Make sure
that the standard has not expired.
Put the correct calibration standard in the vial compartment. Make
sure to invert the standard.
The wrong calibration standard is in the vial compartment. Make sure
that the standard has not expired.
Put the correct calibration standard in the vial compartment.
Examine the verification standard. Calibrate the instrument. Refer to
Calibrate the turbidimeter with StablCal Standards
If verification is not successful after calibration, contact technical
support.
Remove large files from the USB flash drive that use too much space.
Start the instrument update procedure again.
Remove the instrument update files from the USB flash drive. Save
the instrument update files again to the USB flash drive.
Connect the USB flash drive to the instrument. Start the instrument
update procedure again.
Remove the instrument update files from the USB flash drive. Save
the instrument update files again to the USB flash drive.
Connect the USB flash drive to the instrument. Start the instrument
update procedure again.
Contact technical support.
Connect a USB flash drive to the instrument. Make sure that the file
system "FAT32" is installed on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Connect the USB flash drive. Start the instrument update
procedure again.
Connect a USB flash drive to the instrument. Make sure that the file
system "FAT32" is installed on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Look for remaining space on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Connect the USB flash drive to the instrument.
Connect a USB flash drive to the instrument. Make sure that the file
system "FAT32" is installed on the USB flash drive.
Set the power to off, wait 20 seconds and then set the power to on
again. Connect the USB flash drive. Start the instrument update
procedure again.
Solution
on page 13.