5.6 Replace the tubing
Keep water out of the vial compartment or instrument damage will occur. Before the automatic cleaning module is
installed on the instrument, make sure that there are no water leaks. Make sure that all tubing is fully seated.
Make sure that the vial nut is tight.
Replace the tubing when the tubing has a blockage or has damage.
1. Set the flow shut-off valve to off. Install the automatic cleaning module on the service bracket.
Refer to steps 1 to 3 of
2. Replace the tubing.
3. Set the flow shut-off valve to on. Make sure that there are no water leaks. Refer to steps 5B and
6B of
Install the automatic cleaning module
4. Install the automatic cleaning module on the turbidimeter. Refer to step 8B of
cleaning module
on page 7.
Section 6 Replacement parts and accessories
Personal injury hazard. Use of non-approved parts may cause personal injury, damage to the
instrument or equipment malfunction. The replacement parts in this section are approved by the
manufacturer.
Note: Product and Article numbers may vary for some selling regions. Contact the appropriate distributor or refer to
the company website for contact information.
Replacement parts
Description
Seal, process vial
Fiber vial wiper, automatic cleaning module
Silicone vial wiper, automatic cleaning module
Vial with seal, process
Vial replacement tool
Accessories
Description
Micro fiber cloth, vial cleaning
Service bracket
Tubing, inlet and outlet of TU5x00 sc, ¼ in. OD
18 English
N O T I C E
Replace the vial
on page 14.
on page 7.
W A R N I N G
Install the automatic
Item no.
LZY918
LZQ176
LZQ165
LZY834
LZY906
Quantity
Item no.
1
LZY945
1
LZY873
4 m
LZY911