Configuring SonoSite iViz
Configuring SonoSite iViz settings
Configuring preferences
Two new preferences have been added:
Anonymize on USB Share
USB Structured Reports
To configure preferences
1 From the SonoSite iViz Settings screen, tap Preferences
2 From the Preferences screen, the following settings are available:
Units - Select the Height and Weight units used for patient information.
Generic Volume Calcs - Can be measured two different ways, depending on the type of structures
you are measuring. Select either of the following:
Ellipsoid - (default) Length x width x height x ∏/6.
Cuboid - Length x width x height.
Thermal Index - Select any of the following:
TIB - Thermal index in bone
TIC - Thermal index in the cranium
TIS - Thermal index in soft tissue
Auto Save Patient Form - When turned on, this feature automatically saves the patient form
periodically during editing to help prevent the loss of patient information.
Clip Recording - Select either of the following:
Prospective - During a scan, this option records a standard length clip of the scan after tapping
Save.
Retrospective - During a scan, this option records a standard length clip of the scan before tapping
Save.
Clip Length - From the drop-down menu, choose a standard recording length for clips.
Configuring SonoSite iViz
.
3