7. Door adjustment
Under normal circumstance the chamber door lock needs no readjustment. Once steam leaking occurs (the
seal fails), you may use the spanner to turn the door seal thightener.
a.
Open the door first.
b.
Insert the spanner in the gap beneath the plastic cover; use the spanner to lock on the adjusting nut (Fig.
7-2). Turn the nut counter clockwise as the figure below (Fig. 7-1). This will tighten the sealing plate.
c.
Turn the nut until the sealing plate is tight. If the door knob is too tight, you may also turn the nut clockwi-
se to loosen it as long as the seal does not leak.
Caution: Never try to readjust the chamber door while the door is locked.