10.2. Draining of water from the machine
Drain the condenser water collector and water reservoir. Connect the water outlet using the one end of the
tube without a connector and the other end of the tube to a drainage sink. Then tum the draining knob anti-
clock wise to drain the water from the machine (Fig 10-4).
Caution: Before any kind of maintenance turn off the power to the small steam sterilizer and allow it
to cool down. Use only distilled water for sterilizing and cleaning. The instruments being sterilized must be
cleaned. Instruments with oil or impurities must be packed in a sterilization bag.
10.3. Clean and instan the drain filter
Over time small impurities will deposit on the drain filter, and this may affect vacuum and drying. In order to
prolong the life of the drain filter please use the following cleaning procedure:
1. Chamber water filter.
Note: Please check the O ring of Drainning Filter during cleaning. lf this part is broken please replace
it with O ring in the accessories supplied.
54750_AUTOCLAVE_IFU-OK.indd 15
Figure 10-4
Nut
Blue
rubber
2. Back-out the water filter.
3. Take out the filter net, clean it
by soft brush or replace a new one.