a.
Hold the edge of the seal with one hand and insert the screw driver beneath the seal with your other
hand and then pull out the seal slowly.
b.
Once a section of the seal is out, you may pull the whole sea! out slowly with both of your hands. You
can now clean the seal groove and the seal. If the seal is broken or damaged, you must replace it with a
new one.
c.
After the groove and the seal are cleaned, you may remount the seal back in place.
Caution: The seal should be mounted evenly inside the groove with first 4 positioning points in place.
You may use your hand to press the seal in place evenly.
d.
Attention: When inserting the seal, the inner edge might evaginate. You may use a screw driver to press
the edge down in the groove.
Caution: Power the machine off and let it cool down before you change the seal and avoid any burning injuries.
10.7. Cleaning Steam generator
a.
Aim: Clean the steam generator and steam pipe, to avoid
the furring block the pipe.
b.
Description: 20 min cleaning + 10 min drying.
c.
Purpose: to clean the steam generator furring, take it out of
the machine through pre vacuum.
d.
Operation: when the machine is inactivity, press "SELECT"
for 5 seconds, the B&D light and LEAK light will go on at
mean time, it means you chose the cleaning cycle. Close
the door and press "START", it will start to clean.
e.
Cleaning remind: after cleaning, every 100 cycle, B&D light
and LEAK light will going on at the mean time and glitter.
Remind you to clean the steam generator.
f.
Please keep regular cleaning of steam generator to prolong
the service life of the steam generator.
g.
Before executing clean steam generator program, we
suggest to mat a wet towel on the buttom of the chamber
(as shown in below picture), prevent excessive impurities
absorption on the drainning filter, which may influence
pre-vacuum and dry effect.
Friendly remind: Cleaning function will not affect the small
steam sterilizer cycle, only remind message.
10.8. If the machine need to be maintained or the electric
spare parts need to be changed, please cut off the power
first. And ask your dealer or manufacturer to change it.