5. Control panel

Pressure: It displays actual pressure of chamber,
1.
the unit is bar.
Temperature: It displays actual temperature of
2.
chamber.
State displays / Error code displays: It displays
3.
all the states of cycle (Look at the appendix 1).
When it occurs error, it will show the error code,
just according to error code mode.
Caution: lf you meet problems, please con-
tact with dealer or manufacture for more details.
«SELECT» key: Cycle select. According to
4.
different things to be sterilized choose different
temperature you need.
«PACKAGE» key: Package select. You can
5.
choose package or non package according to
the instrument. Non package adopt 3 times pre
vacuum, package adopt B class pre vacuum.
«TEST» key: It used for select the testing cycle.
6.
Press one time,B&D light goes on, means will
run B&D cycle, press again,LEAK light goes one,
means will run LEAK cycle. Press this key for 3s,
two lights goes on at mean time, means to run
cleaning cycle, for cleaning the steam generator.
«DRY» key: This key selects the Fast sterilization
7.
cycle. With unwrapped instruments, you may
use the fast cycle by pressing the DRY key once,
the indication light goes on and it indicates that
machine is in fast cycle mode. Using the Small
Steam Sterilizer made by us, you may still use this selection to shorten the cycle time.
«STEP» key: This is a post cycle selection key. When a sterilize cycle completes,you may press this key
8.
to keep on working or stop working The indication light goes on, and it means that when the cycle is
complete, the heating element will stop before the chamber door is opened. When the indication light is
off, it means that the chamber will be kept warm after completing a cycle and heating up time is shorter
during next cycle.If the user does not start any operation after completing the cycle, the system will enter
into last cycle automatically an hour later.
«START/STOP» key: When you choose the right program, push this button then it starts a cycle. Kee-
9.
ping pressing the key 3 seconds before desiccation phase, the cycle of sterilization will be quitted. Then
it will enter into 3 minutes dryness. In this program if you press the the key again, the program of dryness
will be over at any moment. If you press the key at end of dryness, the cycle will be quitted directly.
10. «WATER» indication light: When the water volume is low, the light is glinting.
11. «READY» indication light: When the light goes on, the machines gets ready for sterilization, and the
operator can press "start" key to start a cycle.
12. «RUNNING» indication light: Working light. It means sterilize program is working.
Caution: In case of faults, please contact your distributor or the manufacture immediately.