15. Troubleshooting

a.
During venting or drying, the thump sounds. It means that the condensed water collector is full. You must
drain the water from The collector.
b.
Hissing from the chamber door (Solving this problem according to item 7 Door Adjustment).
c.
During a cycle, a banging sounds and large steam leaks from the chamber door and displays "E6" code.
Please verify whether the door knob is properly tumed and locked. If the steam leaks at the beginning of
a cycle, you need to change the seal according to instructions at page 18.
d.
When a cycle completes, there are water drops on the instruments. You may correct the problem by:
1.
Please check the drain filter.
2.
Reduce the load of the instrument in the chamber, especially the porous and dressing and other
water absorbent materials.
3.
You may also wait for a while at the end of a cycle "Ed" for better drying result. The small steam
sterilizer will continue heating the chamber until you open the door.

16. Important information

a.
The small steam sterilizer must be installed on a leveled counter top.
b.
When operated ata higher elevation ofmore than 500 meters,The machine must be reprogrammed. Plea-
se get in touch with your distributor for instructions.
c.
Distilled water must be used to prolong the life ofthe machine.
d.
The radiation openings in the machine must not be blocked or covered.
e.
lnstruments should be placed on the tray allowing space between one another and thus allowing steam
to pass through.
f.
Condensed water collector must be emptied from time to time.lt should be emptied when refilling the
water reservoir (Draining of the collector can be followed as per item 10.2).
g.
Chamber door knob must be tumed to its locking position.
h.
Never try to open the chamber door before pressure displays "0.0".
i.
To avoid buming injuries, power off the machine and let it cool down before replacing the seal.
j.
No dragging when the machine is moving.
k.
Power connection must be grounded properly.
l.
Must be provided enough power (1800VA).
m. While the temperature of outside is lower, please pre-heating the machine without instruments for 30
minutes before sterilize.
n.
It is normal ifthe machine alarms when it starts for the first time. lf it alarms again, please contact your
distributor.