9. System setup
a.
Enter function setup mode: Press "STEP" key for 20s when on non-work state (time dispaly "Ld") pressu-
re display "1", enter function setup mode.
b.
Function mode setup: With 7 options, press "SELECT" key, you can select 1-7-1 cycle dispaly. The
definition and display of 7 options as following:
-
Option 1: year, display "yer", press "STEP" or « » key increase, press "DRY" or « » key decrease,
setup numerical value.
-
Option 2: month, display "nth", press "STEP" or « » key increase, press "DRY" or « » key decrea-
se, setup numerical value.
-
Option 3: day, display "day", press "STEP" or « » key increase, press "DRY" or « » key decrease,
setup numerical value.
-
Option 4: hour, display "hor", press "STEP" or « » key increase, press "DRY" or « » key decrease,
setup numerical value.
-
Option 5: minute, display "nin", press "STEP" or « » key increase, press "DRY" or « » key decrea-
se, setup numerical value.
-
Option 6: Printing, it display "Pvt", press "STEP" key then you can choose the cycle between "on"
and "off". "on" stand for "print", "off" stand for "no printing".
-
Option 7: Exit, when "Eit" displays, please press "START/STOP" to save and exit setup mode.
-
Option 8: Sterilization time adjust,display "S-T" when you setup this option, the sterilization time
(SP) win increase at former value.The adjustable range are following 3 kinds. One is increased by 1
minute, second is increased by 2 minutes.Third is increased by 3 minutes.
-
Option 9: temperature data adjust automatically option, display D-P, When you setup this option,
you can adjust the temperature more or less than 2C-3C inorder to meetyuqr request. But we sug-
gest user should setup under guidance of teahnician or not setup.
Press " SELECT" key after setup, you will come back to option 1. At this time, you must press STEP key
3 seconds in option 7. Then you can start the setup of option 8 and option. Otherwise it only can start
cycle between optionland option 7.
Caution: If sorne function setup was changed by pressing the key without intention,please switch off
the power,then restart it.It will exit from the wrong setup.In this case,the function won't be affected by this.
10. Maintenance & Maintain
10.1. How to clean the water tank
Please clean the sterilizer water tank and waste water tank regularly. The process is following:
a.
Empty the tanks water, please refer to item 10.2 draining the water.
b.
Open the tank cover board, you can use screw driver to loosen the five screws just as fig 10-1 indication.
c.
Pull up the water tank cover board by one hand,just, as fig 10-2 indication.
d.
Pull up the tank cover completely and you can disassemble the tank cover. As fig 10-3 indication.
e.
Now you can start to clean the tank, use cotton to dip in sorne alcohol or medical disinfectant to scour
the tank wall, wash it by distilled water, then empty the tank and dry the tank. After it you should take out
the filter net to clean.
f.
After cleaning the tank, assemble the tank cover and tighten the screws.
Figure 10-1
14
54750_AUTOCLAVE_IFU-OK.indd 14
Figure 10-2
Figure 10-3