14. Appendixes
14.1. Work mode chart
Sterilizing
temperature
Sterilizing
pressure
Classes
Pre-heating
(min)
Vaccumize time
(min)
Admission time
(min)
Sterilizing time
(min)
Drying time (min)
Quick drying
time
Total time (min)
14.2. Error code chart
Item
1
2
3
4
5
6
7
8
9
10
11
12
54750_AUTOCLAVE_IFU-OK.indd 19
134ºC
134ºC non
121ºC
package
package
package
134
134
121
2.10
2.10
1.10
B
S
B
0-7
0-7
0-7
10
4
10
5
5
5
6
4
20
10
10
10
1/5
1/5
1/5
38(29/33)
30(21/25)
52(43/47)
Enor Code
Error
E1
Seren fault
E2
Over pressure (>2.6Mba)
Chamer over temperatura (>145
E3
degree)
E4
Keep warm keep pressure faild
E5
Exhaust failed (20S<0.3Mba)
E6
Door opened when working
Steam generator over temperature
E7
(>230 degree)
E8
Pre-heating fault>20min
E9
Fill stern fault>30min
A01
Unsuitable water quality.
A02
Lack of water in the clean water tank
A03
Dirty water tank full
121ºC non
134B+
BD test
package
121
134
134
1.10
2.10
2.10
S
B
B
0-7
0-7
0-7
4
10
10
5
5
5
16
18
3.5
10
10
1
1/5
1/5
—
42(33/37)
50(41/45)
26.5
Description
Check sensor
Check inner temperature sensor
pressure sensor
Check outer temperature sensor
Rerun parameter autoadjust
program
Check air relery valve vacuum valve
Check door switch electrical lock
Check steam generator tempera-
ture sensor
Cheack heating ring or heating
pipes
Check fin steam pipes
The use of water with a conductivity
of more than 15µS/cm can affect
the sterilization process and damage
the sterilizer.
Refill tank with clean water
Empty the dirty water tank
Vacuum
CLEAN
test cycle
test
—
—
—
—
—
—
—
—
4
Clean 20
—
Drying
Keep
pressure
—
10
—
10
—
—
14
30
19