d.
After selecting above setup, put the instruments into the Steam
Sterilizer which need sterilize.
Caution: Instruments that are ready to sterilize should be placed
in the tray to touching one another allowed space for steam to pass
through. lt is suggest that the tray holder be used to put the tray with
instruments inside the chamber to avoid buming injuries.
e.
After the instruments are loaded, you may close and lock the chamber door by tuming the door knob
clockwise to it is locking position. The "ready" indication light goes on. (Fig 6-4, 6-5, 6-6).
Figure 6-4
Caution: The door knob must be tumed to its lock position,otherwise, there might be steam leak during
a cycle and the alarm goes on and displays error code "E6" . Ifthe chamber is in keep warm mode, it is
sometimes difficult to close the chamber door because there is still steam inside. You can keepthe door open
for a while until steam and moisture is vaporized,and the door is easy to close. You may also push the door
inwards with sorne force while tuming the door knob to its locking position. If there is still steam leakafter you
did the above, you may try to read just the door mechanism by a spanner as per item 7. Door adjustrnent.
f.
Press "START/STOP" key, the "ready" light goes off and display
"HE" the machine starts a sterilization cycle. It takes about 30-60
minutes to complete an automatic cycle that includes heating up,
sterilizing and drying. The cycle time depends on the volume of the
instruments loaded, the chamber temperature at the beginning of
the cycle and the cycle program you selected.
g.
When a cycle completes, the display "Ed" the indication light
"RUNNING" Tums off and the indication light "READY" tums on
with sounds. At this time you may then open the chamber door and
unload the instruments which are sterilized. (Fig 6-6) At mean time,
temperature indicate "prt", it means that it is performing print, whe-
ther you have install printer,the process can not be omitted. But you
needn't care about it. It will not effect the process of opening the
door and unloading the instruments. The indication will disappear
when printing is over.
Attention: To ensure small steam sterilizer's on the safe side, When
water level low alann sounds, Please be advised add sorne water first.
When you add the water, please don't forget to let condenser water out.
Caution: Always use the tray holder to load or unload the tray with
instruments in it, in order to avoid burning injuries.