Instructions For Replacing The Suction Arm - B.A. International Optima EOS350 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 97
Extraoral Suction System, Instructions For Use and Installation
7.2 Suction Arms
[Daily Clean]
•Use a spray disinfectant or wipes to clean the systems surface.
•To internally clean the system, turn it on to its lowest suction level, spray hydrogen
peroxide disinfectant based solution into the suction tubes and let the system run for 3
minutes.
•Caution: Do not disassemble the suction arms and soak them in a disinfectant.
[Replacement]
•Turn off the power and remove power cord
from wall.
•Keep your hands dry.
•Wear gloves.
•Loosen up the screws in the first suction tube.
•Remove the entire suction tube and replace it
with a new arm.
•Tighten up the screws.
[Caution]
Replace suction tubes.
Dispose of as medical waste.
16

Publicidad

Tabla de contenido
loading

Tabla de contenido