7. Disengage the suture anchor from the
insertion device:
a. Hold the handle securely in the palm
of the hand and pull back on the finger
grips using two fingers to retract the
suture tensioning mechanism.
b. While holding the suture tensioning
mechanism in the back position with
the finger grips, slowly remove the
insertion device from the insertion site
until the suture tips are visible. The
suture releases and feeds through the
insertion device as it is removed.
Alternatively:
a. Pull back on the finger grips and unwind
the suture with the free hand, then
release the finger grips.
b. Slowly remove the insertion device.
c. As the insertion device is removed,
grasp the suture below the insertion
device using the free hand to verify
suture release from the insertion device.
Note: A gentle back-and-forth twisting
motion can be helpful in removing the
anchor inserter from the curved guide.
CAUTION: Do not use sharp instruments
to manage or control the suture.
8. Discard the flexible insertion device.
Warranty
For single use only. This product is warranted
to be free from defects in material and
workmanship. Do not reuse.
For Further Information
If further information on this product is needed,
contact Smith & Nephew Customer Service at
+1 800 343 5717 in the U.S., or an authorized
representative.