Operating environment

Note the following points so that the instrument will operate correctly and
have a long service life.
Place the instrument firmly on an even surface. Do not push any
objects under the instrument.
The ambient temperature must be 5 to 40 °C (41 to 104 °F).
The relative humidity should be less than 80%; moisture should not
condense on the instrument.
A minimum clearance of 15 cm (5.9 in.) must be ensured above and
on all sides of the device; this allows the air to circulate and
prevents overheating of the electric parts. The dosage arm of the
stepper motor operating the syringe piston must be able to move
freely.
Do not operate or store the instrument in extremely dusty, humid or
wet locations.
Make sure that no liquids enter the instrument and immediately wipe
away any liquids that contact the instrument.
Protect the instrument against vibrations, direct sunlight, corrosive
gases as well as strong magnetic and/or electric fields.
Only use genuine replacement parts and accessories.
The instrument must be serviced by the manufacturer's service
department at the prescribed intervals.
Installation
Installation of the burette
The burette dispenses the titrant, i.e. the 0.01 M I
The burette comprises the syringe, stepper motor and valve. The
stepper motor and valve are already installed in the TitraLab; the syringe
must be installed by the user.
English 8
Warning against system damage
Make sure that the syringe is never operated without the attachment sleeve.
Only use genuine replacement parts and accessories.
The instrument must be serviced by the manufacturer's service department at the
prescribed intervals.
Note: Make sure that you do not misplace the attachment sleeve for the
syringe. It is not provided with a new syringe.
1.
2.
3.
4.
Risk of injury
During the first days of system use, check and tighten the screw joint between the
syringe and valve to avoid leaks and the potential escape of reagents.
solution.
3
N O T I C E
Fully insert the syringe piston into the syringe body.
Slide the attachment sleeve fully onto the syringe so that the
narrow end points toward the piston.
With the piston facing downward, place the syringe into the bracket
such that the attachment sleeve is above the bracket.
With the narrow side facing down, carefully press the attachment
sleeve into the bracket.
Hold the attachment sleeve in the bracket and screw the syringe
onto the valve from below.
Withdraw the piston from the syringe and place it onto the screw in
the dosage arm.
Hold the syringe piston and turn the attachment screw on the
dosage arm into the thread in the piston.
WA R N I N G