pH Calibration

For Display setup, analog outputs and setpoints see the section "General Setup".
pH Two point Calibration
7.0
A
MΩ-cm
25.0
A
°C
Calibrate Sensor
m
Channel A pH
8
Selecting Yes and pressing [ENTER] means a setpoint can be configured for channel a.
Following setpoint Types can be selected:
– High (High value has to be set)
– Low (Low value has to be set)
– Between (High and Low value has to be set)
– Outside (High and Low value has to be set)
After setting the setpoint value(s) a Relay (none, 1, 2) can be configured for
that setpoint. The Relay delay is set to 10 seconds and the Hysteresis is set to 5%. If a
relay is configured as CLEAN relay, it will not be selectable in this menu.
(PATH: Cal)
While in Measurement mode press the [CAL] key. If the display prompts you to enter
the calibration security code, press the
then press the [ENTER] key to confirm the calibration security code.
For Multi-channel devices: Using the
user change the channel to be calibrated.
Press [ENTER] to select the pH calibration sub function.
A flashing "H" (alternating with "A" or "B" to show the channel being calibrated for
multi-channel devices) in the top left hand corner shows the ongoing calibration process.
or
key to set the calibration security code,
m
.
or
key on the "Channel A" field lets the
m
.