Features; Definitions; Instructions; Care Instructions - Stryker SYSTEM G 7302-120-000 Instrucciones De Uso

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 6
7302-120-700 Rev-A

Features

NOTE: The Sterilization Case, 1 Hand Piece is shown for
reference.
C
B
A
A
Handle
B
Latch
C
Bracket
D
Lid
E
Base
F
Perforations (Holes and Slots)

Definitions

The symbols located on the equipment and/or labeling
are defined in this section or in the Symbol Definition
Chart. See the Symbol Definition Chart supplied with the
equipment.
Symbol
Definition
General Warning Sign
Consult instructions for use

Instructions

To remove the lid from the base, pull the two latches up
and away from the base. Unhook the latches from the lid
and remove the lid.
www.stryker.com
To secure the lid to the base, hook the two latches onto
the lid. Push the latches down and toward the base.
D
E

Care Instructions

F
For general processing instructions, see the care
instructions manual supplied with your handpiece.

Maximum Quantity of Devices

WARNING: DO NOT exceed the specified maximum
quantity of devices in a sterilization case. The
maximum quantity of devices is based on Stryker
validation of this equipment. Exceeding the maximum
quantity of devices may prevent proper sterilization
of the equipment and/or result in excessive weight.
Always follow the current local recommendations
and/or regulations governing weights and weight
limits.
Sterilization Case, 1 Hand Piece
Device Type
Attachment
Accessory (Chuck Key)
Transfer Shield
Total of All Devices
Sterilization Case, 2 Hand Pieces
Device Type
Attachment
Accessory (Chuck Key)
Transfer Shield
Total of All Devices
Maximum Quantity
Handpiece
Maximum Quantity
Handpiece
EN
1
5
2
1
9
2
5
2
2
11
3