0210-100-700 Rev-T
Features
A
G
J
I
H
E
A
Applied Part(s) – The tip assembly and handpiece body are the applied
parts as defined by the manufacturer according to the standards listed
under Product Safety Certification in the Specifications section.
B
Full Speed Lock Mechanism – The mechanism is used to hold the
trigger at the high irrigation flow setting. For high flow, squeeze the
trigger and push the button. To unlock, squeeze the trigger again.
C
Suction Clamp – The clamp is used to open and close the suction
tubing pathway to control suction at the handpiece tip assembly.
D
Suction Tubing – The suction tubing provides a suction pathway from
the handpiece tip assembly to the vacuum fluid collection canister
connected to the facility vacuum source.
E
Irrigation Tubing/Power Cord Combination – The irrigation tubing
provides a fluid pathway from the irrigation bag to the handpiece tip
assembly. The power cord conducts electrical energy from the battery
power pack to the handpiece.
F
Trigger – An activated trigger provides irrigation to the surgical/wound
site.
G
Tip Locking Mechanism – The mechanism is used to engage and
release the tip assembly for the handpiece.
H
Battery Power Pack Housing – The battery power pack housing stores
the battery power pack and allows easy removal of the battery pack.
The battery pack provides power to the handpiece. (REF 0210-XXX-000
Series)
I
Irrigation Tube Spike – The spike is used to pierce and connect the
handpiece irrigation tubing to the irrigation bag.
J
Irrigation Tube Spike Cap – The protective spike cap is installed over
the sharp irrigation tube spike to maintain sterility and prevent damage
or injury.
K
Power Cord and Connector – The connector is installed into an
InterPulse Rechargeable Power Pack. (REF 0211-XXX-000)
L
Suction Tube Adapter – The adapter is used for a large diameter port
on a vacuum fluid collection canister and may be removed if necessary.
www.stryker.com
Print Date: Nov 02, 2016 02:57:27 PM
0000004059, Rev. T Effective Date: Sep 26, 2016 2:47:57 PM
B
F
E
J
I
K
E
Definitions
The symbols located on the equipment and/or labeling are defined in this section
or in the Symbol Definition Chart. See the Symbol Definition Chart supplied with
the equipment.
SYMBOL
C
D
Instructions
L
To Assemble the Handpiece
WARNINGS:
• For wound care applications performed outside the sterile field where the
handpiece is used on a single patient repeatedly, ALWAYS disinfect the
D
handpiece, and replace the tip assembly and suction tubing before each use.
See the Cleaning section and the instructions for use supplied with the Wound
Care Tubing Replacement Sets.
• DO NOT reuse the handpiece for surgical applications or applications
requiring a sterile field. See the instructions for use supplied with the Wound
Care Tubing Replacement Sets.
• ALWAYS use aseptic technique when handling the sterile handpiece, tip
assembly, and irrigation tubing.
• ALWAYS use aseptic technique when removing the irrigation tube spike cap.
Failure to comply may cause the irrigation tube spike to become contaminated
before insertion into the irrigation bag.
• DO NOT allow fluid from the irrigation bag to drip onto a battery power pack.
Failure to comply may cause electrical shock.
• DO NOT submerge the battery power pack into liquid or irrigation solution.
• DO NOT use a rechargeable power pack that gives off an odor, generates
heat, becomes discolored, or leaks.
NOTES:
• Make sure a standard IV pole is equipped with an irrigation bag and a fully
charged InterPulse Rechargeable Power Pack, if a power pack is required.
• For optimal performance and maximum run time, use rechargeable power
packs that have been charged within 48 hours.
• Make sure the top of the fluid contained in the irrigation bag is less than 36
inches above the handpiece at all times after connecting the handpiece.
1. If using a handpiece with a rechargeable power pack, install the handpiece
power cord plug into the receptacle of the InterPulse Rechargeable Power
Pack REF 0211-022-000 (figure 1). See the instructions for use supplied with
the InterPulse Power Pack Charger.
2. Using aseptic technique, remove the irrigation tube spike cap from the
handpiece irrigation tubing and install the irrigation tube spike into the irrigation
bag aseptic peel port.
DEFINITION
General warning sign
Battery lid removal tabs
(REF 0210-XXX-000 Series only)
Type B Applied Part
Direct current (DC)
EN