10.1 | Technical specifi cations for TENS mode
Waveform
Pulse width:
Pulse rate:
Burst:
Normal:
Pulse width modulation:
Pulse rate modulation:
10.2 | Technical specifi cations for EMS mode
Pulse Width:
Puls Rate
Ramp up time:
On time:
Ramp down time:
Off time:
11 | Explanation of symbols on unit
LOT
SN
IP22
IP22
Symbol for CONSULT Operation
Guide
Symbol for TYPE BF APPLIED
PARTS
single alternate square pulse wave
from 200 to 250µS depends on the diff erent programs
from 2 to 80Hz according to diff erent programs
TENS 3: There is one pulse wave includes 7 pulses which is 100
Hz per 2 Hz output.
TENS 1: The pulse rate and pulse width output will be 80 Hz /
200 µS;
TENS 2: The pulse rate and pulse width output will be 2 Hz /
250 µS;
TENS 4: The pulse width is automatically varied in a cycle time.
The pulse width output is 2Hz/250µS in the fi rst 3 second
and 80 Hz/200 µS in another 3 seconds. Thus run in
cycles.
TENS 5: The pulse rate is automatically varied in a cycle time. The
pulse rate is increased from 2 Hz to 80 Hz in setting cycle
time 5 s, keeps 1second and then decreased from 80 Hz
to 1 Hz in nest setting cycle time 5 s and keep 1 s, thus
run in cycles. In this program, pulse rate (2 to 80 Hz),
pulse width 200 µS
250 microseconds
EMS 1: 50 Hz
EMS 2: 20 Hz
EMS 3: 30~80 Hz
EMS 4: 30~45 Hz
EMS 5: 30~45 Hz
Output amplitude rising edge time within a period 2s± 30%
Output amplitude keeping time within a period 5 ± 30%
Output amplitude trailing edge time within a period 2s ± 30%
Output amplitude zero output time within a period 10s ± 30%
0483
Not suitable for children under
3 years!
Waste electrical products should
not be disposed of with household
waste
Symbol for MANUFACTURER
COMPILES WITH MDD93/42/EEC
REQUIREMENTS