Universal commands
Command # 0 Read Unique Identifier
request data bytes
response data bytes
response codes
command #1 read Primary Variable
request data bytes
response data bytes
response codes
command #2 read current And Percent Of range
request data bytes
response data bytes
response codes
command #3 read current And Four dynamic Variables
request data bytes
response data bytes
response codes
command #6 Write Polling Address
request data bytes
response data bytes
response codes
none
#0 – 254 (fixed)
#1 - Manufacturer Id = 255 (general)
#2 - Manufacturer device type = 32 for Ultrasonic Transducer
#3 - number of Preambles
#4 - Univ cmd rev
#5 - trans spec rev
#6 - Soft Rev (37 for 3.7)
#7 - Hard rev (12 for 1.2)
#8 - Flags
#9 to #11 - device Id number (24-bit unsigned int)
#0 - No Command-Specific Errors
none
#0 - PV Units code
#1 to #4 - Primary Variable
#0 - No Command-Specific Errors
none
#0 to #3 - P.V. current [mA]
#4 to #7 - P.V. Percent of Range [%]
#0 - No Command-Specific Errors
none
#0 to #3 - P.V. current [mA]
#4 to #7 - P.V. Percent of Range [%]
#0 - No Command-Specific Errors
#0 - Polling Address of device
#0 - Polling Address of device
#0 - No Command-Specific Errors
#2 - Invalid Selection (Address > 15)
#5 - too Few data bytes received
#32 - busy