sYsteM setUP

Press E and Up arrow key until "SYSTEM SETUP" menu appears on the display screen
in order to enter system setup. Use up and down arrow keys for running between
menus. Press E key instantly in order to change the setting inside the menu. Press
Left arrow key instantly in order to exit from the menu or use BACK option.

1. current (analog Outlet setting)

It is the menu where 4 mA and 20 mA calibration of analog output is made. 4mA
and 20mA was entered as initial value. the value can be increased and decreased by
pressing up and down arrow keys after pressing E key. It is saved by long pressing to
E key again or by pressing E key after save option is selected.
Note: This setting is made in factory environment. It should not be changed unless
any technical problem is encountered.

2. temp set (temperature Offset)

It is used for the cases when sensor internal temperature value is not equal with the
outer environment. For compensation, 0Oc was entered as default value in menu
where you will enter offset value. You can change the value by pressing up and
down arrow keys after pressing E key. It is saved by long pressing to E key again or by
pressing E key after save option is selected.
• It should be used in order to perform an accurate measurement for the cases
where sensor internal temperature value is not equal with the outer environment
temperature value. the sensor's measurement for the temperature difference
between internal environment and the outer environment should be accepted as
the temperature offset value.

3. simulation

this menu is used to test the device to where it is connected. select one of the
LEVEL, cUrrEntor tEMP options by using arrow keys and enter by pressing E key.
change the desired value by pressing E key again and using arrow keys.

4. system Default

It is used to restore factory settings of the device. "system setup" is entered by
pressing E key. E key is pressed after "DEFAULTS LOAD NO" option is made as
"dEFAULts LOAd YEs" by up arrow key. It is asked if you are sure about this process
as "DEFAULTS SURE?". E key is pressed if this process is desired to be performed and
left arrow key is pressed if the process is desired to be cancelled.
16
 Orion Echo Ultrasonic Level Probe