Descargar Imprimir esta página

Orion Echo Manual De Instrucciones página 34

Publicidad

It defines shape of depot or silo if volume is selected
40410
sHAPE
as measuring type. 0= rectangle, 1= cylinder, 2=
horizontal cylinder 3= sphere
temperature measuring unit 13= degree, 14=
40412
tEMP UnIt
Fahrenheit
temperature value to be added to or to be
extracted from the value measured in temperature
measurement. Minus value is extracted from the
tEMP
measured one, plus value is added to measured
40414
OFFsEt
one.
Its unit defined with tEMP UnIt. Its decimal value
is always 1.
It is the value to be taken for 4 mA output for
purpose of scaling 4..20mA output. Its unit is
40416
I sPAn LOW
tYPE_UnIt (40400), and the decimal value is MAIn_
VALUE_dEcIMALs (40104).
It is the value to be taken for 20 mA output for
purpose of scaling 4..20mA output. Its unit is
40418
I sPAn HIGH
tYPE_UnIt (40400), and the decimal value is MAIn_
VALUE_dEcIMALs (40104).
It is decimal value for dimensions. (dIMX, dIMY etc)
(table 3)
dIM_
Use only maximum value mentioned in Table 3. The
40420
dEcIMALs
values except this is brought to maximum decimal
value for the unit selected automatically during
writing process by the devices.
40422
dIM_UnIt
It defines the unit to be used for dimensions.
zero point for main degree distance measurement
(as from end of the device)
zErO
40424
OFFsEt
Its unit is dIM_UnIt, its decimal value is dIM_
dEcIMALs
zero point for main degree level measurement (as
40426
LEVEL stArt
from end of the device). Its unit is dIM_UnIt, its
decimal value is dIM_dEcIMALs
main value volume measuring silo/depot depth (as
40428
dIM_X
from end of the device). Its unit is dIM_UnIt, its
decimal value is dIM_dEcIMALs
rectangular shaped silo/depot width Y size for main
40430
dIM_Y
value volume measuring. Its unit is dIM_UnIt, its
decimal value is dIM_dEcIMALs
rectangular shaped silo/depot width z size for main
40432
dIM_z
value volume measuring. Its unit is dIM_UnIt, its
decimal value is dIM_dEcIMALs
Horizontal/ Vertical cylinder shaped silo/depot
diameter d size for main value volume measuring.
40434
dIM_d
Width of Parshall flume. Its unit is dIM_UnIt, its
decimal value is dIM_dEcIMALs
Horizontal cylinder shaped silo/depot length L size
40436
dIM_L
for main value volume measuring. Its unit is dIM_
UnIt, its decimal value is dIM_dEcIMALs
It is used to reset Panshal flume meter. to reset the
PArsHAL
meter; firstly 12345 is written while the value in the
FLUME
40500
register was 0. then 0 is written again after it is seen
MEtEr
(40124) from relevant register that the meter is reset.
rEsEt
34
 Orion Echo Ultrasonic Level Probe
Long
4
0..2
Long
4
13 – 14
-200...
degree
Long
4
200
Fahrenheit
M, cm, inch,
Feet, M
Long
4
-
Liter, Us
Gallon
M, cm, inch
Feet, M
Long
4
Liter, Us
Gallon
Long
4
0..3
Long
4
1..4
M, cm, inch
Long
4
-
M, cm, inch
Long
4
-
M, cm, inch
Long
4
-
M, cm, inch
Long
4
-
M, cm, inch
Long
4
-
M, cm, inch
Long
4
-
M, cm, inch
Long
4
-
Long
4
r/W
-
-
,
3
,
3
-
Feet
Feet
Feet
Feet
Feet
Feet
Feet
-

Publicidad