SECTION 6—MAINTENANCE
SECTION 6—MAINTENANCE
NOTE: The Invacare concentrators are specifically designed to
minimize routine preventive maintenance. In places with high dust
or soot levels, maintenance may need to be performed more often.
Only authorized or factory-trained personnel should perform
preventive maintenance on the concentrator. Power should be
disconnected before beginning preventive maintenance on the
Invacare concentrator.
NOTE: At a minimum, preventive maintenance MUST be
performed according to the maintenance record guidelines. Refer to
Preventive Maintenance Record Invacare Platinum Oxygen
Concentrator on page 30.
Routine Maintenance
NOTE: The following routine maintenance should be performed
more frequently. See recommended intervals in each section and
preventive maintenance record.
Cleaning the Cabinet Filter
NOTE: For this procedure, refer to FIGURE 6.1 on page 29.
CAUTION
DO NOT operate the concentrator without the filters
installed.
NOTE: There are two cabinet filters. One filter located on each side
of the cabinet.
1. Remove each filter and clean at least once a week
depending on environmental conditions.
NOTE: Environmental conditions that may require more frequent
cleaning of the filters include but are not limited to: high dust,
smoking, air pollutants, etc.
Platinum
28
Part No 1141492
™