Flowrate
NOTE: For this procedure, refer to FIGURE 5.6.
1. Turn the flowrate knob to the setting prescribed by your
physician or therapist.
DO NOT change the L/min. setting on the flowmeter
unless a change has been prescribed by your physician or
therapist.
NOTE: To properly read the flowmeter, locate the prescribed flowrate
line on the flowmeter. Next, turn the flow knob until the ball rises to
the line. Now, center the ball on the L/min. line prescribed.
2. If the flowrate on the flowmeter ever falls below 0.5 L/min.
for more than about one minute, the LOW FLOW alarm
will be triggered. This is a rapid beeping of the audible
alarm. Check your tubing or accessories for blocked or
kinked tubing or a defective humidifier bottle. After rated
flow is restored, more than 0.5 L/min., the LOW FLOW
audible alarm will go off.
NOTE: The use of some accessories such as the pediatric flowstand
and the HomeFill compressor will deactivate the Low Flow Alarm.
Part No 1141492
SECTION 5—OPERATING INSTRUCTIONS
WARNING
FIGURE 5.6 Flowrate
25
Flow Knob
Ball
Platinum