Invacare Platinum IRC5LXAW Manual De Instrucciones página 33

Ocultar thumbs Ver también para Platinum IRC5LXAW:
Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 105
SYMPTOM:
YELLOW or RED
light illuminated.
Alarm:
Continuous
On red light only.
* Only applies to
SensO
models.
2
GREEN light
YELLOW Light
Flashing.
Alarm:
RAPID
Beep....Beep...
Beep...Beep
Part No 1141492
SECTION 7—TROUBLESHOOTING GUIDE
PROBABLE CAUSE:
1. Low oxygen purity.*
2. Kinked or blocked
tubing, cannula or
humidifier.*
3. Flowmeter set at 0.5
L/min. or less.*
4. Unit overheating due
to blocked air intake.
5. Insufficient power at
outlet.
6. Internal repairs
required.
1. Internal repairs
required.
Low Flow Alarm
1a. Kinked or blocked
tubing, cannula or
humidifier.
1b. Flowmeter set at 0.5
L/min. or less.
33
SOLUTION:
1. Clean or Replace filters.
2. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turn power OFF
for 60 seconds and then
turn power back ON.
3. Check flowmeter is set to
1.0 L/min. or more. (Refer
to Recommended
Guidelines for Optimum
Performance on page 20.
4.a. Remove and clean
cabinet filters.
4b. Move concentrator at
least 7.5 cm (3 in) from
walls, draperies, and
furniture.
5. DO NOT use extension
cords. Move to another
electrical outlet or circuit.
6. Call Service Provider.
1. Call Service Provider.
1a. Inspect for kinks or
blockages. Correct, clean or
replace item. Once
corrected, turnpower OFF
for 60 seconds and then
turn power back ON.
1b. Reset flowmeter to
prescribed flowrate.
Flowrates less than1L/min.
are not recommended.
Platinum

Publicidad

Tabla de contenido
loading

Este manual también es adecuado para:

Platinum irc5lxo2awPlatinum irc5lxo2awqPlatinum irc9lxo2awq

Tabla de contenido