Commissioning
!
For evaluation purposes, use the table under "Two-point calibration".
5.
Switch back to the measuring mode with <M>.
12.5 Configuring
!
Note!
You can leave the configuration menu at any time with <M> . The parameters that have already
been changed are stored.
1.
Switch off the measuring instrument.
2.
Press <M> and <
3.
Press <CAL> to choose what should be displayed during the calibration.
You can choose from:
– Buffer target value/pH
– Current electrode voltage mV
4.
Press <RUN/ENTER>.
The measuring instrument automatically switches to the measuring mode.
12.6 Resetting to factory settings
1.
Press <RUN/ENTER> and <CAL> simultaneously.
2.
Press <RUN/ENTER> again.
All the settings are reset to the factory settings and the instrument switches to the measur-
ing mode.
The following factory settings exist:
Parameter
Measuring mode
Asymmetry
Slope
Type of calibration
Resolution of the pH display
26
Note!
I
> simultaneously.
Value
pH
0 mV
-59.16 mV/pH
AutoCal TEC
0.01
CPM280
Endress+Hauser