• Do not wrap the cuff over jacket or sweater sleeve or measurement cannot be done.
• Remove tight-fitting clothing from your left arm.
• Do not in any way twist the arm cuff.
• Do not inflate the monitor's cuff when it is not wrapped around the arm.
• Do not attempt to disassemble or change any parts of the monitor including the cuff.
• Do not drop the product or put it through strong impact.
• The device is not supposed to be used if your arm has any wound or injury.
Applying the arm cuff
1. Slide the end of arm cuff furthest from the tube through the metal ring to a loop.
The smooth cloth should be on the inside of the cuff.
Pile side material
Velcro
Smooth
cloth
Metal ring
Fig. 4
2. If the cuff is located correctly, the Velcro will be on the outside of the cuff and the metal ring
will not touch the skin. (Fig. 4)
3. Put your left arm through the cuff loop. The bottom of the cuff should be approximately
(2~3 cm) above elbow. The tube should lie over the brachial artery on the inside of the arm.
(Fig. 5)
4. Pull the cuff so that the top and bottom edges are tightened around your arm. (Fig. 6)
5. When the cuff is positioned properly, press the Velcro firmly against the pile side of the cuff.
6. This cuff is suitable for use if the <<index>> mark falls within the <<ok range>> marked by
two arrows when the cuff is tightened around your arm. (Fig. 7)
7. Sit on a chair and place your arm on the table so that the cuff is at the same level as your
heart. (Fig. 8)
Fig. 5
OK
OK
Fig. 7
Left arm
Main
arteries
Fig. 8
Fig. 6
5