13.5 Initiating blood processing

13.5.1 Blood processing
Warning: Do not apply pressure to the holding bag, as this can lead to air embolism.
Note: Respond to prompts that appear during operation. See Section 18.4 and Section 19.4 for a list of prompts that may appear
during operation.
1. The message Machine stopped. Press
2. Press
to activate the self-start mode. The message Self-start enabled. Press
Note: The device automatically starts when it detects approximately 800 mL of blood in the reservoir for at least 5 s.
3. If it is desired to begin the fill before 800 mL is collected, press .
4. The default wash volume is 250 mL. The wash volume level can be adjusted during a case; adjustments are effective
immediately. If a different wash volume is desired, select one from the Wash volume menu on the Main screen.
English
M967226A006 1B Medtronic Confidential
Response
Note: There is a feature on the manifold cover
that pushes the tubing into the air detector when
closing the cover.
6. Rotate the manifold cover latch until the lock indi-
cators are aligned.
1. Spike the middle manifold tubing into the saline
bag or bags.
Note: If using only 1 saline bag, confirm that the
clamp on the unused line is closed.
2. Open the saline clamp or clamps.
3. Connect the left-hand tube of the manifold to the
blue-capped step-down connector at the bottom
of the reservoir, and then open the clamp on the
step-down connector.
4. Press .
to resume.
If the device detects air during the prime or wash pha-
ses, this message appears.
1. Confirm that the saline bags are mounted and
spiked. Mount and spike the saline bags if nec-
essary.
2. If the previous step does not resolve the mes-
sage, reset the alert multiple times.
3. If previous steps do not resolve the message,
ensure that the tubing is correctly inserted into the
air detector.
4. If the previous steps do not resolve the message,
clean the air detector and tubing with water, and
then dry.
5. If the previous steps do not resolve the message,
use a different device if available. Contact your
Medtronic service representative.
Warning: Close all clamps and use caution if
transferring the disposables to a new device.
1. Open the manifold cover.
to resume.
2. Place the manifold into the recessed cavity.
3. Place the 3 manifold tubing lines into the tubing
guides, and ensure that each tube is fully seated
in each guide.
4. Place the single tubing line of the manifold into
the air detector, and ensure that it is fully seated.
Note: There is a feature on the manifold cover
that pushes the tubing into the air detector when
closing the cover.
5. Close the manifold cover.
6. Rotate the manifold cover latch until the lock indi-
cators are aligned.
7. Press .
to enable self-start. appears.
Audible alert
signal
None
Low
Medium
to manually start. appears.
30
502431-132
Visual alert
signal
None
Low
Medium
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018

Publicidad

Tabla de contenido
loading

Tabla de contenido