Size: 7 x 9 in (178 x 229 mm)
Prompt
Machine stopped. Press
start.
Self-start enabled. Press
start.
Vacuum pressure detected at startup.
Make sure the suction line is not clam-
ped. Press
to resume.
18.4 Normal operating prompts, audible and visual alert signals
Normal prompts may appear during device operation or startup, and they typically can be resolved by making a quick adjustment
of the device or the disposable kit.
Prompt
Air in reservoir line. Press
select Return, Concentrate, or Wash.
Air in holding bag line. Bowl is not full.
Press
to try again or select Return or
Wash.
Kit is not installed. Lock the bowl cover.
Kit is not installed. Lock the manifold
cover. Press
to resume.
M967226A006 1B Medtronic Confidential
Response
to enable self-
This message appears after the disposable kit is
installed. Enable self-start by pressing . The
device automatically starts when it detects approx-
imately 800 mL of blood in the reservoir for at least
5 s.
to manually
Press
to cause a manual start.
This message appears if existing vacuum pressure
is detected in the reservoir when vacuum is turned
on at device startup.
1. Confirm that vacuum and suction lines are not
clamped or immersed in liquid.
2. To continue, press .
3. Press
to restore vacuum.
Note: If prompt returns, the device can still be
used if an external regulated vacuum source is
connected to the reservoir. Contact your
Medtronic service representative.
Response
to try again or
The Fill phase emptied all the contents of the res-
ervoir.
1. If possible, remove air from the reservoir line.
Press
to continue filling.
2. Depending on the clinical use condition, select
Return, Concentrate, or Wash.
Warning: Choosing to wash a partially filled
bowl may result in low blood quality and low red
cell volume.
This message appears after selecting Concen-
trate, and after the red blood cells from the holding
bag are consumed before the bowl is full.
1. Unclamp the holding bag line if necessary.
2. Press and hold
3. If the previous steps do not resolve the mes-
sage, select Return to return the blood to the
reservoir or Wash to wash the contents of the
bowl.
4. The centrifuge continues to spin for 1 min, and
blood is returned to the reservoir if no selection
is made.
Warning: Choosing to wash a partially filled
bowl may result in low blood quality and low red
cell volume.
A sensor has detected that the bowl cover is not
locked.
1. Place the bowl in the centrifuge chamber with
the yellow-capped waste line outlet tubing fac-
ing the waste bag. Refer to the markings on the
top panel.
2. Align the centrifuge markings with the vertical
assemblies on the bowl.
3. To lock the bowl, simultaneously press down
on it and turn it clockwise until a click is heard.
A sensor has detected that the manifold cover is not
a
locked.
1. Open the manifold cover.
2. Place the manifold into the recessed cavity.
Audible alert
signal
None
None
Medium
Audible alert
signal
Low
Low
to advance blood to the bowl.
None
Medium
43
502431-132
Visual alert
signal
None
None
Medium
Visual alert
signal
Low
Low
None
Medium
English