Notifications - Medtronic autoLog IQ Manual Del Usuario

Tabla de contenido

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

Size: 7 x 9 in (178 x 229 mm)
Figure 38. Service and maintenance screen
1 Current device notifications
2 Current device information such as software operating system, graphical user interface, and system controller
versions; serial number; and device, centrifuge, roller pump, and vacuum pump run times
3 Logged device errors
4 Allows the user to calibrate the touchscreen
5 Allows the user to perform the centrifuge cleaning operation
6 Allows the service technician to perform authorized software updates
7 Export service information (not shown): Allows the user to export notifications, device information, and error logs

19.4 Notifications

Notifications provide information about device conditions that are important but do not stop device operation. The presence of a
notification is indicated by a yellow outline on the
Service and maintenance screen (Figure 38). The following information describes the appearance and behavior of device
notifications:
• A new notification, or a postponed notification that has reappeared, causes yellow outlines to appear in the screen areas
described previously.
• Pressing the Reset alert field allows the user to postpone a yellow-outlined notification. Postponing the notification causes the
yellow outlines to disappear from the screen areas described previously. The following reset options are available:
– Minutes: Select this option, and then use
– Cases: Select this option, and then use
– Forever: For applicable notifications, select this option and the notification will not reappear.
• Notifications disappear entirely once the underlying condition is corrected. For example, the notification Max vacuum is still
set. Adjust if necessary. disappears after the vacuum setting is changed.
• Active notifications, whether yellow-outlined or not, can be accessed by pressing Notifications from the Service and
maintenance screen.
• If there are active notifications, and if there are no higher-priority prompts also active, the notifications will continuously appear
(scroll) in the Prompts area of the Main screen.
The following table contains a list of device notifications:
Table 2. Notifications, audible and visual alert signals
Notification
Maintenance due in __ days.
The real-time clock failed. See error log
for details. Contact Medtronic for service.
Max vacuum is still set. Adjust if neces-
sary.
Emergency wash is still set. Adjust if nec-
essary.
Case records will not be stored unless a
password is set in Settings.
M967226A006 1B Medtronic Confidential
and the Reset alert field (Figure 39), and on the Notifications field of the
and
to set the number of minutes before a notification reappears.
and
to set the number of cases before a notification reappears.
Description
The device begins counting down when mainte-
nance is due in 30 days, and updates the counter
daily. Contact your Medtronic service representa-
tive to perform maintenance on the device.
Case records will not contain an accurate time
stamp. Contact your Medtronic service representa-
tive.
Max vacuum has been set for 10 min or more. If
Max vacuum is not desired, change the vacuum
setting as shown in Section 13.3.
Emergency wash has been set for 10 min or more.
If Emergency wash is not desired, change the
Wash volume setting as shown in Section 13.5.1.
Set the password, in order to store case records as
shown in Section 12.4.
49
Audible alert
Visual alert
signal
signal
None
Low
None
Low
None
Low
None
Low
None
Low
English
Composed: 2018-06-08 06:06:55
PSJ - Size-selectable package manual 02-MAY-2018
502431-132

Publicidad

Tabla de contenido
loading

Tabla de contenido