Prompt
Optics obstructed. Empty the bowl or
clean the optics sensor. Press
resume.
Air in saline line. Press
Emptying time was too short. Check the
holding bag line for air. Press
resume.
Emptying time was too long. Check the
holding bag line for occlusions. Press
to resume.
M967226A006 1B Medtronic Confidential
Response
9. If the message is not resolved after the previ-
ous steps, and if the bowl was not replaced,
then replace the bowl.
10. If the message is still not resolved, use another
device if available. Contact your Medtronic
service representative.
Warning: Close all clamps and use caution if
transferring the disposables to a new device.
The level sensor detects blood at the top of the bowl
to
when there should be no blood in the bowl.
1. Examine the bowl for blood.
2. If there is blood in the bowl, manually return the
blood to the reservoir using the FCT key (Fig-
ure 21).
3. If the previous step does not resolve the mes-
sage, unlock the bowl.
4. If clots are observed in the bowl, discard and
replace the wash kit.
5. If clots are observed inside the reservoir filter,
prime the reservoir with anticoagulant.
6. Agitate the reservoir vigorously.
7. If clots are observed outside the reservoir filter,
replace the reservoir.
8. If the previous steps do not resolve the mes-
sage, document all fluid volumes.
9. Clean and dry the optic sensors using an
approved cleaning solution (Chapter 16).
10. If the previous steps do not resolve the mes-
sage, use another device if available. Contact
your Medtronic service representative.
Warning: Close all clamps and use caution if
transferring the disposables to a new device.
to resume.
This message appears if the device detects air dur-
ing the prime or wash phases.
1. Confirm that the saline bags are mounted and
spiked. Mount and spike the saline bags if nec-
essary.
2. If the previous step does not resolve the mes-
sage, reset the alert as needed.
3. If the previous steps do not resolve the mes-
sage, ensure that the tubing is correctly inser-
ted into the air detector.
4. Clean the air detector and tubing with water,
and then dry.
5. If the previous steps do not resolve the mes-
sage, use another device if available. Contact
your Medtronic service representative.
Warning: Close all clamps and use caution if
transferring the disposables to a new device.
Examine the disposables for air leaks.
to
1. Confirm that there is flow to the holding bag.
2. If no flow is observed, remove the centrifuge
bowl and examine it for clots.
Warning: Do not invert the bowl when exam-
ining for clots, as this could lead to fluid leak-
age.
3. If clots are observed in the bowl, discard and
replace the wash kit.
4. If no clots are observed, clean the air detector
and tubing with water, and then dry.
This message appears if the holding bag line is
clamped and the device is unable to move blood
volume from the bowl, or if the air detector has failed.
Audible alert
signal
Low
Low
Low
Low
45
502431-132
Visual alert
signal
Low
Low
Low
Low
English