Turbo-V 6000 Controller User Manual / 87-900-855-01
Connection P1 Logic Input Interconnections
DESCRIPTION
The output Voltage will be present upon the programmed condition delay YES
or delay NO (see the cycle diagram in the following pages).
If YES is selected, R2 is off and the output is zero over all run up time, then:
a.
If running speed > speed threshold
R2 = OFF
b.
If running speed < speed threshold
R2 = ON
If NO is selected:
a.
If running speed > speed threshold
R2 = OFF
b.
If running speed < speed threshold
R2 = OFF
FAULT signal 24 V, 60 mA, optically
negative).
The output Voltage will be present when a fault condition is displayed on the
front panel display.
Analog output voltage (0÷10 V) of pump speed (pin 7 positive, pin 14 negative).
Technical Information
isolated output (pin 13 positive. pin 6