Descargar Imprimir esta página

Phase 3 "Recovery; Important - CareFusion Vyntus WALK Instrucciones De Uso

Publicidad

Idiomas disponibles
  • ES

Idiomas disponibles

  • ESPAÑOL, página 127

Phase 3 "Recovery"

Version 05.00 • 06-06-2017
All manuals and user guides at all-guides.com
The recovery phase will last 10 minutes. During the recovery phase, the SpO2 and
heart rate continue to be measured. Tap <Autofill> repeatedly to update the
current measurement values in the
Systolic and Diastolic blood pressure values can be entered manually.
IMPORTANT: The distance covered in the final lap of the test phase needs to be
FINAL LAP
entered in the
To monitor the patients recovery, touch <Values>. Heart rate and SpO2 values
versus time are continuously displayed in a graphical representation.
Tap on <Next> to start the result phase.
SPOT CHECK DATA
Tap on <Rating> to open the
RPE Scale.
Transfer the results of the patient's
subjective evaluation to the scale
(see Phase 1 "Baseline").
field.
Touch <Set> to close the window.
The values in the
DATA
Instructions for Use Vyntus WALK
table.
SPOT CHECK
field will be updated.
Page 31/332

Publicidad

loading

Este manual también es adecuado para:

Wristox 3150