NOTE: The device also features a thermal fuse that turns off the nebulizer in the event of
overheating. Should this occur, proceed as follows:
• Switch off the appliance
• Unplug the power cord from the outlet.
• Wait for a minimum of 30 minutes until the device has completely cooled down.
• Plug the power cord back into the outlet.
• Then switch the device on again, and observe whether it makes any abnormal noises. If
there are no unusual noises and smells, you can use the device again. Otherwise, contact
your customer service team. (Please see warranty for service contact)
7.

Changing the Filter

DANGER:
• Always unplug this product immediately after use and make sure it is not acces-
sible to children.
CAUTION:
• Do not operate the nebulizer without a filter.
• Check the filter at regular intervals (at the latest after every 10th use). If the filter
is very contaminated or blocked, replace the used filter. If the filter has become
wet, it must also be replaced with a new one.
• Do not try to clean and re-use the used filter!
To replace the filter, proceed as follows:
• First switch off the unit and unplug it.
• Allow the unit to cool down.
Pull the filter cap [6] forward and turn it.
1
• NOTE: If the filter remains in the unit when the cap is removed, remove the filter
from the unit using tweezers or the like.
Replace the filter cap [6] containing the new filter.
2
Check that the filter cap is firmly in place.
3