Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain
a device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to
maintain a device may compromise the clinical condition or safety of users and/or their caregivers. Do not
perform service and maintenance while the user is in the product.
1st Day
•
Read the User Guide carefully before use and store
it in a safe place for future reference
Before every use
•
Disinfect the product. Please check the disinfection
section in this User Guide for more information
After every use
•
Wipe the product dry.
Daily
•
Use a cloth with warm water and a mild detergent
to clean the product
•
Visual inspection. Repair or replace any damaged
or worn-out parts
•
Check all fittings and buckles for signs of worn
out parts
•
Check that all castors are moving freely and that all
wheel locks/brakes are easily activated
•
Ensure that all fasteners are in place and fitted
correctly
Monthly
•
All nuts and bolts on this product must be checked
and tightened to avoid unnecessary failures
Annually
•
Inspect the frame for cracks or signs for worn-out
parts and carry out annual service. Never use a
product which appears to be substandard or faulty
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to wipe
off any residue and dirt from the product, using a
cloth with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
Warewashing machine
This product can be washed at 60° using a mild
detergent for 10 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
This product can be washed at 85° using a mild
detergent for 3 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
Before washing
Remove the following items and clean them
separately: Cushioning + Accessories
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months if the product is
used very often - daily use for more than 5 times),
every time the product is reissued for use and after
a long term storage (more than 4 months). The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 8 years
if all maintenance and servicing is carried out in
www.etac.com
accordance with the manufacturer's instructions
and demonstrably recorded. The device may be
reissued within its lifetime
Service check list:
Check and adjust the following functions as per
their intended use. Repair or replace any damaged
or worn-out parts. Please check the:
•
Accessories can be inserted and locked in the
fittings in the top of the back, using the knob and
handles mounted in the fittings.
•
Seat is adjustable and can be locked in position.
•
Fittings and buckles for signs of worn out parts
•
Product for cracks or signs for worn-out parts
•
All labels on the product are intact
Materials
The following surface treatment processes were
used to protect against corrosion:
•
Lacquered surfaces = Polyester powder coating or
ED-coating
•
Non-lacquered aluminium parts = Anodized coating
•
Non-lacquered steel surfaces = Galvanized
23