R82 Swan Curo Guia Del Usuario página 23

Tabla de contenido

Publicidad

Service information
Maintenance of a medical device is wholly the responsibility of the owner of that device. Failure to maintain
a device in accordance with the instructions may invalidate the device's warranty. Furthermore, failure to
maintain a device may compromise the clinical condition or safety of users and/or their caregivers. Do not
perform service and maintenance while the user is in the product.
1st Day
Read the User Guide carefully before use and store
it in a safe place for future reference
Before every use
Disinfect the product. Please check the disinfection
section in this User Guide for more information
After every use
Wipe the product dry.
Daily
Use a cloth with warm water and a mild detergent
to clean the product
Visual inspection. Repair or replace any damaged
or worn-out parts
Check all fittings and buckles for signs of worn
out parts
Check that all castors are moving freely and that all
wheel locks/brakes are easily activated
Ensure that all fasteners are in place and fitted
correctly
Monthly
All nuts and bolts on this product must be checked
and tightened to avoid unnecessary failures
Annually
Inspect the frame for cracks or signs for worn-out
parts and carry out annual service. Never use a
product which appears to be substandard or faulty
Disinfection
The product can be disinfected with a 70%
disinfectant IPA solution. It is recommended to wipe
off any residue and dirt from the product, using a
cloth with warm water and a mild detergent/soap
without chlorine and let it dry before disinfection.
Warewashing machine
This product can be washed at 60° using a mild
detergent for 10 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
This product can be washed at 85° using a mild
detergent for 3 minutes in a washing machine
designed for medical devices. Use the machine's
drying function to dry the product.
Before washing
Remove the following items and clean them
separately: Cushioning + Accessories
Service interval
A detailed inspection of the product must be made
every 12 months (every 6 months if the product is
used very often - daily use for more than 5 times),
every time the product is reissued for use and after
a long term storage (more than 4 months). The
inspection should be carried out by a technically
competent person who has been trained in the use
of the product.
The lifetime of this product in normal use is 8 years
if all maintenance and servicing is carried out in
www.etac.com
accordance with the manufacturer's instructions
and demonstrably recorded. The device may be
reissued within its lifetime
Service check list:
Check and adjust the following functions as per
their intended use. Repair or replace any damaged
or worn-out parts. Please check the:
Accessories can be inserted and locked in the
fittings in the top of the back, using the knob and
handles mounted in the fittings.
Seat is adjustable and can be locked in position.
Fittings and buckles for signs of worn out parts
Product for cracks or signs for worn-out parts
All labels on the product are intact
Materials
The following surface treatment processes were
used to protect against corrosion:
Lacquered surfaces = Polyester powder coating or
ED-coating
Non-lacquered aluminium parts = Anodized coating
Non-lacquered steel surfaces = Galvanized
23
EN

Publicidad

Tabla de contenido
loading

Tabla de contenido