The Hydrostatic Testing Date
Aluminum cylinders MUST be tested every five years.
DO NOT fill cylinders that have not been tested in the
past five years. Otherwise, serious personal injury may
result. For replacement contact your supplier.
External Examination
1. Examine the outside of the cylinder for the following
conditions, and replace the cylinder in the event of:
• Dents or surface damage
• Arc burns
• Oil or grease
• Any other signs of damage which may cause the
cylinder to be unacceptable or unsafe for use.
2. Inspect the cylinder for evidence of fire or thermal
damage. Such evidence includes charring or blistering of
the paint or other protective coating or heat sensitive
indicator. If fire or thermal damage is found, replace the
cylinder.
3. Inspect the cylinder/regulator assembly for the following,
and replace if found:
• Debris, oil or grease
• Noticeable signs of damage
• Signs of corrosion inside the valve
• Signs of excessive heat or fire damage
Part No 1145805