SECTION 5—INDICATOR LIGHTS
INDICATOR LIGHT
COLOUR
None
O
Below Normal -
2
YELLOW
Filling - GREEN
Full - GREEN
Part No 1145805
COMPRESSOR
STATUS
Compressor is Off.
Compressor is On and
warming up (0-3
minutes after power is
On).
O
from concentrator
2
is <90% (anytime after
three minute warmup).
Compressor is On and
cylinder is filling. O
from concentrator is
>90% (any time after
three minute warmup).
Compressor is On.
Cylinder has finished
filling (full) -remove
cylinder (FULL light is
On in >10 mins.)
or
Cylinder is not
attached properly -
check connection
(FULL light is On in
<10 mins.).