Turning the Compressor On

NOTE: For this procedure, refer to FIGURE 4.5 on page 28.
1. Make sure the concentrator is On. Refer to the
concentrator Owner's Manual.
The concentrator flow rate to the patient MUST be set at
a maximum of 2.5 L/min. or less for Platinum 5 models
and 5 L/min. for Platinum 9 models. Otherwise, the
oxygen level to the compressor will be of <90% and the
compressor will not fill the cylinder.
2. Set the concentrator flow rate to the patient to 2.5 L/min. or
less for Platinum 5 and 5 L/min. for Platinum 9 oxygen
concentrator. Refer to the concentrator Owner's Manual.
3. Make sure the cylinder is connected to the compressor.
Refer to Connecting the Cylinder to the Compressor on
page 22.
4. Push the power switch on the control panel to the On ( )
position.
5. Examine the control panel. The following sequence of
events should occur:
A. 0-3 minutes - The O
will light up while the compressor warms up.
B. After three minutes - The filling (GREEN) light will be
on while the cylinder is filling.
C. The full (GREEN) light will be on when the cylinder is
finished filling. Proceed to step 6.
NOTE: The O
oxygen level from the concentrator has not yet reached or drops
below 90%. The compressor will begin or resume filling when the
Part No 1145805
CAUTION
below normal (YELLOW) light may come on if the
2
below normal (YELLOW) light
2
27
HomeFill
II Compressor
®