OBJ_DOKU-26720-006.fm Seite 19 Freitag, 7. April 2017 12:57 12
5
Compatible devices
The products are only intended for use with the following Erbe
electrosurgical units:
Epilation handle
20190-008
20190-064
6
How to use
6.1
Only use sterile products
Clean, disinfect and sterilize the product before using it for the
first time and before every subsequent use.
6.2
Check the product
Before each use, check the product including all the insulation
for signs of physical damage!
If damaged, do not use this product!
6.3
Assembling the product
Insert the HF electrode firmly into the handle as far as it goes!
6.4
Connect up the product and select unit settings
Plug the connecting cable into the monopolar socket of the elec-
trosurgical unit.
Only select unit settings that do not exceed the maximum elec-
trical capacity of the product (V
units, read the "Accessories" chapter.
Electrosurgical unit
Erbe T series
Erbe ACC / ICC / VIO
max). For Erbe electrosurgical
p
19