6.5

Using the product

Footswitches are used to activate the product.
Press the yellow pedal to activate cutting mode. Press the blue
pedal to activate coagulation mode. Make sure that the electro-
surgical unit responds correctly.
7

Cleaning, disinfection, sterilization

7.1

Safety instructions

Erbe Elektromedizin recommends machine cleaning/
disinfection. Because it is much less effective, manual cleaning
is not recommended.
Not suitable for cleaning using an ultrasonic bath.
Never use sharp objects for cleaning.
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Validated procedures
in overview".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
7.2

Reprocessing limitation

Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
20