Calibration Procedure _ Cal 4 - Baxtran IKP Serie Manual De Usuario

Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

IKP model
Print
ALL.2
Print
oFF
Note: To use complete print function, M+ setting in CAL 1 must be set to Add Pr.
4. Peripheral setting
DISPLAY
rENOtE
LP50
rENOtE
PC

CALIBRATION PROCEDURE _ CAL 4

Please remove the JP3 jumper before you start the calibration process.
In Calibration menu, press
and press
to confirm.
1. SEL maximum capacity
DISPLAY
Kg
Lb
LoAd
LoAd
03
06
LoAd
LoAd
06
12
LoAd
LoAd
15
30
LoAd
LoAd
30
60
LoAd
LoAd
50
100
Note: The capacity will display in kg or lb according to the preset weight unit in CAL 1.
2. Zero point calibration
DISPLAY
XXXXXX
0
Pnt.0
Make sure the weighing pan is empty. When the AD value becomes stable, press
3. Setup calibration weight
DISPLAY
XXXXXX
XXXXX
Pnt.CAL
Put on poise of the calibration weight. Use the number keys to input the calibration weight in gram.
Press
to print single weighing result.
Press
in
mode to print accumulated weighing result and total accumulation.
Print Net wt when the scale is stable and weight >0.
Print function disabled.
DESCRIPTIONS
Connect to LP-50 thermal printer.
Connect to PC for data transmission.
to toggle among the menu options. Use the numeric keys to input data
DESCRIPTIONS
Max. capacity displayed varies with preset wt unit
Set maximum capacity as 3Kg/6lb
Set maximum capacity as 6Kg/12lb
Set maximum capacity as 15Kg/30lb
Set maximum capacity as 30Kg/60lb
Set maximum capacity as 50Kg/100lb
DESCRIPTION
The first row shows AD value, the second row shows "0" and the third row shows "Pnt.0".
DESCRIPTION
The first row shows AD value, the second row is the calibration weight (Unit: gram or
lb/1000) and the third row shows "Pnt.CAL".
The default calibration weight is 1/3 of scale capacity. Use the number keys to change the
desired calibration weight.