Data Management
Data Management
The TrueCPR device can store compression data for three 60-minute sessions,
or up to six sessions totaling 180 minutes. When all available memory has been
used, the data from the oldest use is overwritten automatically.
Data can be transferred to a computer with TrueCPR device-compatible software
via USB connection. Event reports can be printed directly from the software. For
information on TrueCPR device reports and full event review functionality,
contact your local Physio-Control representative.
WARNING
SHOCK HAZARD: Do not use the USB connector if the TrueCPR device
is damaged.
POSSIBLE ELECTRICAL INTERFERENCE: Using cables or accessories
that are not specified for use with the TrueCPR device may result in increased
emissions or decreased immunity from electromagnetic or radio frequency
interference (RFI) for this device. This could affect the performance of the
TrueCPR device or equipment in close proximity. Use only cables and
accessories specified in these instructions for use.
To initiate data transfer:
1 Connect a USB cable between the TrueCPR device and a computer with
compatible software.
Note:
Use only a USB cable that meets the following specifications: USB 2.0
A-Male to Mini-B 5-pin Male, 28/24 AWG cable with ferrite core (gold plated),
1 m (3 ft).
2 Turn on the TrueCPR device. If the software on the computer does not start
automatically, open the program manually.
3 Follow the instructions in the software on the computer to select the TrueCPR
device and initiate data transfer. Data transfer may take up to 5 minutes if
memory is full. The following screens may appear on the TrueCPR device:
Data transfer in
progress.
12
All manuals and user guides at all-guides.com
Data transfer successful.
Data transfer failed.
TrueCPR Device Instructions for Use