Product Specifications - Physio Control TrueCPR Instrucciones De Uso

Ocultar thumbs Ver también para TrueCPR:
Tabla de contenido

Publicidad

Idiomas disponibles

Idiomas disponibles

Product Specifications

Product Specifications
SPECIFICATION
Power Requirements
Operating Time
Operating Temperature Device can be used at 0° to 40°C (32° to 104°F).
Shelf Life
Storage Temperature
Relative Humidity
Altitude
Object and Water
Resistance
Weight
Ventilation Prompts
Compression Depth
Metronome Rate
Model
Safety Classifications
CATEGORY
Protection Against Electric Shock
Protection Against Solid Object and
Liquid Ingress
Mode of Operation
* Type BF defibrillation proof applied parts are the underside of the chest pad and the top
of the back pad. See page 15 for symbol illustration.
16
All manuals and user guides at all-guides.com
DESCRIPTION
Batteries: 2 Duracell nonrechargeable DL123 cells
180 minutes at room temperature with new batteries
Applied part temperature may reach 43°C (109°F) during
extended use at ambient temperature of 40°C (104°F).
Provides 30 minutes of CPR feedback after 24 months without
use at 25°C (77°F), starting with new batteries installed.
-30° to 70°C (-22° to 158°F)
5 to 90%, non-condensing
-382 to 4,572 m (-1,253 to 15,000 ft)
IP55 per IEC 60529 and EN 1789
Less than 0.75 kg (1.65 lb) with batteries installed
2 ventilation prompts every 30 compressions in No Airway mode.
No ventilation prompts in Airway mode.
Target depth range of 5 to 6 cm (2 to 2.5 in)
±
104.4
1 compressions per minute, consistent with AHA and
ERC guidelines.
80596-000001
CLASSIFICATION
Internally powered ME equipment.
Type BF defibrillation proof applied parts.*
IP55
Continuous operation
TrueCPR Device Instructions for Use

Publicidad

Tabla de contenido
loading

Tabla de contenido