Troubleshooting Tips for Data Management
OBSERVATION
Feedback screen appears on
TrueCPR device and
metronome sounds.
Data Transfer Failed screen
appears.
Cleaning the Device
The TrueCPR device should be cleaned and inspected after each use.
1 Clean the device with a sponge or cloth that has been dampened with one of
the cleaning agents shown below:
•
Quaternary ammonium compounds
•
Isopropyl alcohol
•
Hydrogen peroxide, 3% solution
•
Sodium dichloroisocyanurate (NaDCC), 3000 ppm solution
•
Chlorine bleach, 1:10 dilution with water
2 Inspect the device and cable for wear or damage. Remove from service if
damaged.
3 After cleaning, wipe off solution with a clean, damp cloth.
IMPORTANT!
Do not immerse or soak the chest pad or back pad.
Storing the Device
Place chest pad onto back pad. Store in carrying bag so that battery readiness
indicator is visible.
CAUTION
POSSIBLE DEVICE DAMAGE: If the TrueCPR device will be stored for an
extended period of time, remove the batteries to prevent possible damage
due to battery leakage.
©2012 Physio-Control, Inc.
All manuals and user guides at all-guides.com
CORRECTIVE ACTION
• Confirm USB cable is properly connected.
• Confirm computer is on.
• Replace USB cable.
• Use recommended USB cable (see Step 1
on page 12).
• Turn TrueCPR device off and check
connections. Then turn device on and
attempt data transfer again.
• If problem persists, contact Technical
Support.
English
13