16. Inspect the housing of the transmitter for any signs of cracking, flaking, or damage. If these signs
are present, the transmitter must be discarded according to local regulations for battery disposal
(non-incineration).
WARNING: Do not use the device if there is any cracking, flaking, or damage to the
housing. Cracking, flaking, or damage to the housing are signs of deterioration.
Deterioration of the housing can affect the ability to properly clean the transmitter, and
result in serious injury. Call a local Medtronic support representative and discard the
device according to local regulations for battery disposal (non-incineration), or contact
a healthcare professional for disposal information.
17. Discard the used gloves and thoroughly wash hands with soap and water.
Removing adhesive residue
Follow these instructions if there is adhesive residue present on the transmitter.
Use cotton swabs and a medical adhesive remover such as Detachol™*, a mineral spirit, to remove
adhesive residue.
Note: During testing, Detachol™* was used to remove the adhesive residue from the transmitter.
Detachol™* is recommended for use but may not be available in all countries.
To remove adhesive residue:
1. Make sure the tester is attached to the transmitter.
2. Soak a cotton swab in the medical adhesive remover.
3. Hold the tester and gently rub the adhesive remover on the transmitter until the residue is
removed.